Perbandingan Hasil High Sensitivity Troponin I Point of Care Testing dengan Immunology Analyzer Laboratorium Sentral pada Sindrom Koroner Akut
| dc.contributor.author | Putri, Andina Ayu | |
| dc.date.accessioned | 2026-09-24T17:23:08Z | |
| dc.date.issued | 2024-08-20 | |
| dc.description | Prof. dr. Rismawati Yaswir, Sp.PK, Subsp. K.V. (K), Subsp. E.M. (K); Dr. dr. Dwi Yulia, Sp.PK, Subsp. Onk.K. (K) | |
| dc.description.abstract | Latar Belakang : Kebutuhan biomarker sensitif dan spesifik dalam mendiagnosis sindrom koroner akut (SKA) mendorong perkembangan metode immunoassay pemeriksaan troponin (cTn). High sensitivity cTn (hs-cTn) merupakan biomarker yang direkomendasikan untuk identifikasi kerusakan miokard. Metode point of care testing (POCT) hs-cTn dapat mengurangi turn around times (TAT) dalam pengambilan keputusan cepat di Instalasi Gawat Darurat (IGD). Penelitian bertujuan untuk mengetahui perbandingan hasil hs-cTn I POCT metode Flourescence Immunoassay (FIA) sampel whole blood dan plasma dengan immunology analyzer sampel serum metode Enzyme-Linked Fluorescent Assay (ELFA). Metode : Penelitian analitik dengan rancangan potong lintang dilakukan kepada 31 pasien SKA dari IGD RS Dr. M. Djamil Padang pada bulan November 2023 – April 2024. Subjek penelitian dapat dilakukan pengambilan sampel whole blood EDTA. Analisis menggunakan Passing Bablok Regression dan Bland and Altman Plot. Hasil : Median hasil pemeriksaan hs-cTn I menggunakan immunology analyzer, POCT whole blood dan plasma adalah 87,70, 90,50, dan 115,0 ng/L. Perbandingan hasil hs-cTn I whole blood POCT dengan immunology analyzer didapatkan r=0,985 (p<0,001) dan persamaan y=0,55+0,88x menggunakan Passing Bablok Regression, rerata perbedaan dua pengukuran didapatkan 27,14 ng/L (95%CI; -115,84-170,12; p<0,001) dan nilai bias 8,51% (95%CI; -69,73%-86,75%) menggunakan Bland and Altman Plot. Perbandingan hasil hs-cTn I plasma POCT dengan immunology analyzer didapatkan r=0,980 (p<0,001) dan persamaan y=0,11+0,86x menggunakan Passing Bablok Regression, rerata perbedaan dua pengukuran didapatkan 27,26 ng/L (95%CI; -140,49-195,01; p<0,001) dan nilai bias 10,65% (95%CI; -94,21%-115,51%) menggunakan Bland and Altman Plot. Simpulan: Ditemukan perbedaan bermakna secara signifikan antara hasil hs-cTn I POCT dengan immunology analyzer laboratorium sentral dengan nilai bias 8,51% untuk sampel whole blood dan 10,65% untuk sample plasma | |
| dc.identifier.uri | https://ailib.my.id/handle/123456789/4010 | |
| dc.language.iso | id | |
| dc.publisher | Universitas Andalas | |
| dc.subject | Sindrom Koroner Akut | |
| dc.subject | POCT | |
| dc.subject | whole blood | |
| dc.subject | high sensitivity troponin I | |
| dc.subject.classification | R Medicine (General) | |
| dc.title | Perbandingan Hasil High Sensitivity Troponin I Point of Care Testing dengan Immunology Analyzer Laboratorium Sentral pada Sindrom Koroner Akut | |
| dc.type | Thesis | |
| dcterms.dateSubmitted | 2024-08-21 | |
| dcterms.references | Ahsan, H. 2022. Monoplex and multiplex immunoassays: approval, advancements, and alternatives. Comparative Clinical Pathology, 31(2), 333-345. Bergmark, B. A., Mathenge, N., Merlini, P. A., Lawrence-Wright, M. B., & Giugliano, R. P. 2022. Acute coronary syndromes. The Lancet, 399(10332), 1347-1358. Bruinen, A. L., Frenk, L. D., de Theije, F., Kemper, D. W., Janssen, M. J., Rahel, B. M et al. 2022. Point-of-care high-sensitivity troponin-I analysis in capillary blood for acute coronary syndrome diagnostics. Clinical Chemistry and Laboratory Medicine (CCLM), 60(10), 1669-1674. Byrne, R. A., Rossello, X., Coughlan, J. J., Barbato, E., Berry, C., Chieffo, A et al. 2023. 2023 ESC Guidelines for the management of acute coronary syndromes: Developed by the task force on the management of acute coronary syndromes of the European Society of Cardiology (ESC). European heart journal, 44(38), 3720-3826. Boeddinghaus, J., Nestelberger, T., Koechlin, L., Wussler, D., Lopez-Ayala, P., Walter, J. E et al. 2020. Early diagnosis of myocardial infarction with point-of-care high-sensitivity cardiac troponin I. Journal of the American college of cardiology, 75(10), 1111-1124. Christenson, R. H., Frenk, L. D., de Graaf, H. J., van Domburg, T. S., Wijnands, F. P., Foolen et al. 2022. Point-of-care: roadmap for analytical characterization and validation of a high-sensitivity cardiac troponin I assay in plasma and whole blood matrices. The Journal of Applied Laboratory Medicine, 7(4), 971-988. Clerico, A., Zaninotto, M., Passino, C., Padoan, A., Migliardi, M., & Plebani, M. 2021. High-sensitivity methods for cardiac troponins: The mission is not over yet. Advances in clinical chemistry, 103, 215-252. Clerico, A., Zaninotto, M., Aimo, A., Dittadi, R., Carraro, P., Rampoldi, E et al. 2023. High-sensitivity Point of Care Testing (POCT) methods for Cardiac Troponins: analytical features and clinical relevance. biochimica clinica, 1. Collinson, P., Aakre, K. M., Saenger, A., Body, R., Hammarsten, O., Jaffe, A. S et al. 2023. Cardiac troponin measurement at the point of care: educational recommendations on analytical and clinical aspects by the IFCC Committee on Clinical Applications of Cardiac Bio-Markers (IFCC C-CB). Clinical Chemistry and Laboratory Medicine (CCLM), 61(6), 989-998. Cullen, L., Collinson, P. O., & Giannitsis, E. 2022. Point-of-care testing with high-sensitivity cardiac troponin assays: the challenges and opportunities. Emergency Medicine Journal, 39(11), 861-866. Denessen, E. J., Nass, S. I., Bekers, O., Vroemen, W. H., & Mingels, A. M. 2023. Circulating forms of cardiac troponin: a review with implications for clinical practice. Journal of Laboratory and Precision Medicine, 8. Eggers, K. M., Hammarsten, O., & Lindahl, B. 2023. Differences between high-sensitivity cardiac troponin T and I in stable populations: underlying causes and clinical implications. Clinical Chemistry and Laboratory Medicine (CCLM), 61(3), 380-387. Fan, W., Wei, C., Liu, Y., Sun, Q., Tian, Y., Wang, X et al. 2022. The prognostic value of hematologic inflammatory markers in patients with acute coronary syndrome undergoing percutaneous coronary intervention. Clinical and Applied Thrombosis/Hemostasis, 28, 10760296221146183. Febriana, S., Nurulita, A., & Bahrun, U. 2016. Penilaian Uji Troponin I dengan Point of Care Testing. Indonesian Journal of Clinical Pathology and Medical Laboratory, 22(2), 114-118. Fu, Y., Liu, T., Zhang, Z., Li, H., Li, W., & Huang, M. 2022. The crosstalk fluorescence spectroscopy analysis principle and an accurate fluorescence quantitative method for multi-composition fluorescence substances. Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy, 280, 121472. Giannitsis E & Katus HA. 2018. Biomarker in Acute Coronary Syndromes In The ESC Textbook of Intensive and Acute Cardiovascular Care, 2th Ed, Editors : Tubaro M and Vranckx P, Oxford, p : 314-320. Hang, KX., Schulz, K., Sandoval, Y., Smith, S. W., Saenger, A. K., & Apple, F. S. 2023. Analytical performance comparing siemens whole blood point of care Atellica VTLi to the central laboratory plasma Atellica IM high-sensitivity cardiac troponin I assays. Clinical Biochemistry, 114, 79-85. Karłowicz, JD., Kupczyńska, K., Michalski, B., Kapłon-Cieślicka, A., Darocha, S., Dobrowolski, P et al., 2021. Fourth universal definition of myocardial infarction. Selected messages from the European Society of Cardiology document and lessons learned from the new guidelines on ST-segment elevation myocardial infarction and non-ST-segment elevation-acute coronary syndrome. Cardiology Journal, 28(2), 195-201. Kluger, N. J., & Legget, M. E. 2022. Emerging Biomarkers in Acute Coronary Syndromes–A Pathophysiologic Perspective. Heart, Lung and Circulation, 31(6), 779-786. Leite, L., Matos, P., Leon-Justel, A., Espirito-Santo, C., Rodríguez-Padial, L., Rodrigues, F et al. 2022. High sensitivity troponins: A potential biomarkers of cardiovascular risk for primary prevention. Frontiers in Cardiovascular Medicine, 9, 1054959. López, E. R, Barallat, J., Cserkóová, A., Gálvez-Montón, C., Jaffe, A. S., Januzzi, J. L., & Bayes-Genis, A. 2021. Pre-analytical considerations in biomarker research: Focus on cardiovascular disease. Clinical Chemistry and Laboratory Medicine (CCLM), 59(11), 1747-1760. Mazhar, N., Khan, U. H., Mand, B., Rehman, A., Mubeen, M., Malik, M. H. B. A., & Hussain, S. A. 2022. Frequency of risk factors of coronary heart diseases in patients with acute coronary syndrome and its comparison in male and female. Moazenzadeh, M., Hedayati, M., & Rashidinejad, H. 2023. Gender and age-based differences in risk factors and symptoms of acute coronary syndrome in a sample of iranian patients. Journal of Kerman University of Medical Sciences, 30(2), 64-71. Mohammadzadeh, S., Matani, N., Soleimani, N., & Bazrafshan Drissi, H. 2022. Comparison of Point‐of‐Care and Highly Sensitive Laboratory Troponin Testing in Patients Suspicious of Acute Myocardial Infarction and Its Efficacy in Clinical Outcome. Cardiology Research and Practice, 2022(1), 6914979. Nardin, M., Verdoia, M., Laera, N., Cao, D., & De Luca, G. 2023. New Insights into Pathophysiology and New Risk Factors for ACS. Journal of Clinical Medicine, 12(8), 2883. Omran, H., Deutsch, M. A., Groezinger, E., Zittermann, A., Renner, A., Neumann, J. T et al. 2022. High-sensitivity cardiac troponin I after coronary artery bypass grafting for post-operative decision-making. European heart journal, 43(25), 2388-2403. Radha, R., & Al-Sayah, M. H. 2021. Development of liposome-based immunoassay for the detection of cardiac troponin I. Molecules, 26(22), 6988. Roth, G. A., Mensah, G. A., Johnson, C. O., Addolorato, G., Ammirati, E., Baddour, L. M., ... & GBD-NHLBI-JACC Global Burden of Cardiovascular Diseases Writing Group. 2020. Global burden of cardiovascular diseases and risk factors, 1990–2019: update from the GBD 2019 study. Journal of the American college of cardiology, 76(25), 2982-3021. Sandoval, Y., Apple, F. S., Mahler, S. A., Body, R., Collinson, P. O., Jaffe, A. S., & International Federation of Clinical Chemistry and Laboratory Medicine Committee on the Clinical Application of Cardiac Biomarkers. 2022. High-sensitivity cardiac troponin and the 2021 AHA/ACC/ASE/CHEST/SAEM/SCCT/SCMR guidelines for the evaluation and diagnosis of acute chest pain. Circulation, 146(7), 569-581. Sorensen, N. A., Neumann, J. T., Ojeda, F., Giannitsis, E., Spanuth, E., Blankenberg, S et al. 2019. Diagnostic evaluation of a high-sensitivity troponin I point-of-care assay. Clinical Chemistry, 65(12), 1592-1601. Riskesdas. 2018. Laporan Kinerja Direktorat Pencegahan Dan Pengendalian Penyakit Tidak Menular. Kementrian Kesehatan Republik Indonesia. Subandriyo B. 2020. Analisis Korelasi Dan Regresi. Badan Pusat Statistik. Theofilis, P., Oikonomou, E., Chasikidis, C., Tsioufis, K., & Tousoulis, D. 2023. Pathophysiology of Acute Coronary Syndromes—Diagnostic and Treatment Considerations. Life, 13(7), 1543. Thulin, I. V. L., Myrmel, G. M. S., Jordalen, S. M. F., Collinson, P., & Aakre, K. M. 2023. Point-of-care high-sensitivity troponin assays: advancements, clinical applications, and implementation considerations. Journal of Laboratory and Precision Medicine, 8. Thupakula, S., Nimmala, S. S. R., Ravula, H., Chekuri, S., & Padiya, R. 2022. Emerging biomarkers for the detection of cardiovascular diseases. The Egyptian Heart Journal, 74(1), 77. Tobing, E. R., Nugraha, J., & Amminuddin, M. 2017. DIAGNOSTIC CONCORDANCE BETWEEN NEXT-GENERATION AND HIGH SENSITIVE TROPONIN-I IN ANGINA PECTORIS PATIENTS. INDONESIAN JOURNAL OF CLINICAL PATHOLOGY AND MEDICAL LABORATORY, 24(1), 64-69. Triage MeterPro High Sensitivity Troponin I Package Insert, Quidel. 2022. Triage MeterPro Manual Book. 2022. Tsui, A. K., Lyon, M. E., van Diepen, S., Goudreau, B. L., Thomas, D., Higgins, T et al. 2019. Analytical concordance of diverse point-of-care and central laboratory troponin I assays. The Journal of Applied Laboratory Medicine, 3(5), 764-774. Wang, Y., Yang, Y., Chen, C., Wang, S., Wang, H., Jing, W et al. 2020. One-step digital immunoassay for rapid and sensitive detection of cardiac troponin I. ACS sensors, 5(4), 1126-1131. Yuan, D., Chu, J., Qian, J., Lin, H., Zhu, G., Chen, F., & Liu, X. 2023. New Concepts on the Pathophysiology of Acute Coronary Syndrome. Reviews in Cardiovascular Medicine, 24(4), 112. Zhang, R., Hong, Y., Shi, J., Zhao, R., Song, Y., Li, Z et al. 2021. Analytical characterization and clinical performance evaluation of a new point-of-care testing system for high-sensitivity cardiac troponin I assay. Annals of Clinical Biochemistry, 58(6), 579-585. | |
| local.eprints.dir | disk0/00/47/70/26 | |
| local.eprints.division | SPPK2024 | |
| local.eprints.id | 477026 | |
| local.eprints.type | thesis/3Magister | |
| local.eprints.uri | http://scholar.unand.ac.id/477026/ | |
| local.fakultas | Fakultas Kedokteran | |
| local.identifier.nim | 2050307302 | |
| local.prodi | Sp-1 Patologi Klinis | |
| local.thesis.level | Karya Spesialis |