Dosimetric Comparison of Primo Monte Carlo Simulation and Eclipse Treatment Planning System for Glioblastoma Radiotherapy Treatment Planning
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Andalas University
Abstract
Accurate radiotherapy planning is crucial to optimize the dose to the tumour while
minimizing exposure to healthy organs. The Eclipse Treatment Planning System
(TPS) with the Analytical Anisotropic Algorithm (AAA) is widely used clinically,
yet it has limitations in heterogeneous regions and high-dose gradient areas. This
study aims to compare the dose distribution calculated by Eclipse TPS (AAA) with
Monte Carlo (MC) simulations using PRIMO software in glioblastoma patients
treated with Intensity Modulated Radiation Therapy (IMRT). This study utilized
DICOM data from four glioblastoma patients planned with Eclipse TPS using 6
MV photon beams. The process began with the modelling and validation of the
Clinac CX LINAC in PRIMO against measured data, which was successfully
achieved with a gamma passing rate of 98.9% using 2%/2 mm criteria. Due to the
unavailability of the MLC80 type used in Eclipse TPS within PRIMO, the treatment
plan parameters from Eclipse were manually re-entered with MLC120 substitution
for the re-simulation. The evaluation was performed by analyzing the dose
distribution, Dose Volume Histogram (DVH), Homogeneity Index (HI), and
Conformity Index (CI). The results showed that the HI values obtained from
PRIMO and Eclipse TPS were in relatively good agreement, whereas the CI values
from PRIMO were lower than those from Eclipse TPS due to differences in dose
calculation algorithms and the use of MLC120 as a substitute for MLC80. PRIMO
simulations also revealed that several organs at risk, particularly the brainstem and
optic nerves, received doses approaching or exceeding tolerance limits, thus
providing additional information for treatment plan evaluation.
Keywords: Analytical Anisotropic Algorithm, glioblastoma, IMRT, Monte Carlo,
PRIMO.
Description
Rico Adrial, S.Si., M.Si ; M. Arif Efendi, S.Si., M.Si., Ph.D