UPT Perpustakaan Universitas Andalas
Repositori PerpustakaanUniversitas Andalas
 

Stress Distribution of the Femur in Pre-Total Hip Arthroplasty (THA) Using the Direct Anterior Approach (DAA) With Finite Element Method

dc.contributor.authorS, Riyura Aqila Zahra
dc.date.accessioned2026-09-24T21:34:39Z
dc.date.issued2026-08-24
dc.descriptionProf. Dr. Eng. Meifal Rusli; Hendery Dahlan, Ph.D
dc.description.abstractThe DAA-THA procedure requires specific lower limb positioning combined with mechanical traction using a traction table, which may introduce complex loading conditions on the femur and surrounding structures, potentially leading to stress concentrations that increase the risk of intraoperative fractures. This study aims to determine the stress distribution in the femur and hip joint during DAA procedures under traction table loading conditions using the finite element method. A three-dimensional model of the hip joint was developed from the LibHip database, and six limb position configurations were simulated under fixed distal femur displacement in ANSYS. The simulation results show that limb position significantly affects the reaction force and von Mises stress required to achieve a given magnitude of hip joint distraction. The configuration of 0° hyperextension and 30° abduction was found to be the most mechanically favorable, producing the lowest reaction force (146 N) and the lowest maximum stress (36.51 MPa), while still achieving a joint space widening (5.4 mm) close to the clinically targeted value (5.5 mm). The maximum von Mises stress was consistently located at the distal femur and femoral neck rather than at the hip joint itself, as the frictionless cartilage-to-cartilage contact ceases to transmit force once joint separation occurs. Across all simulated configurations, the maximum stress obtained (up to 62.08 MPa) remained below the reported ultimate tensile strength of cortical bone (approximately 120–150 MPa), suggesting that axial traction alone, within the range of forces and positions considered, is unlikely to be the primary mechanical cause of intraoperative femoral fracture. The findings are anticipated to provide a biomechanical basis for optimizing surgical techniques in DAA-THA and serve as a reference for improving patient safety during orthopedic procedures.
dc.identifier.urihttps://ailib.my.id/handle/123456789/51510
dc.language.isoid
dc.publisherUniversitas Andalas
dc.subjectTotal Hip Arthroplasty (THA)
dc.subjectDirect Anterior Approach (DAA)
dc.subjectTraction Table
dc.subjectFinite Element Analysis (FEA)
dc.subjectStress Distribution
dc.subjectHip Joint Mechanics
dc.subject.classificationTJ Mechanical engineering and machinery
dc.titleStress Distribution of the Femur in Pre-Total Hip Arthroplasty (THA) Using the Direct Anterior Approach (DAA) With Finite Element Method
dc.typeThesis
dcterms.dateSubmitted2026-08-28
dcterms.references[1] M. J. Lespasio et al., “Hip Osteoarthritis: A Primer,” 2018. doi: 10.7812/TPP/17-084. [2] D. Wernly, J. Wegrzyn, G. Lallemand, J. Mahlouly, C. Tissot, and A. Antoniadis, “Total hip arthroplasty through the direct anterior approach with and without the use of a traction table: a matched-control, retrospective, single-surgeon study,” J. Orthop. Surg. Res., vol. 16, no. 1, Dec. 2021, doi: 10.1186/s13018-020-02184-6. [3] R. Supra, R. Supra, and D. K Agrawal, “Surgical Approaches in Total Hip Arthroplasty,” Journal of Orthopaedics and Sports Medicine, vol. 05, no. 02, 2023, doi: 10.26502/josm.511500106. [4] F. Lenze et al., “Minimally Invasive Total Hip Arthroplasty: A Comparison of Restoring Hip Biomechanics with and without a Traction Table,” In Vivo (Brooklyn)., vol. 36, no. 1, pp. 424–429, Feb. 2022, doi: 10.21873/INVIVO.12720. [5] A. S. McGee, S. R. Huntley, Z. L. Littlefield, H. M. McKissack, A. Shah, and S. M. Naranje, “Hip external rotation and extension angles for exposure and preparation of the proximal femur in direct anterior total hip arthroplasty: A cadaveric study,” J. Clin. Orthop. Trauma, vol. 11, pp. S62–S65, Feb. 2020, doi: 10.1016/j.jcot.2019.07.019. [6] A. Driesman and C. C. Yang, “Clinical outcomes of DAA and related techniques in hip arthroplasty,” Dec. 01, 2023, BioMed Central Ltd. doi: 10.1186/s42836-023-00198-z. [7] S. Ishii, T. Baba, K. Hayashi, Y. Homma, O. Mutou, and M. Ishijima, “Total hip arthroplasty via the direct anterior approach using a conventional traction table and fluoroscopy: A safe and cost-effective technique,” SICOT. J., vol. 10, 2024, doi: 10.1051/sicotj/2024045. [8] N. Ramadanov et al., “Indirect comparisons of traction table versus standard table in total hip arthroplasty through direct anterior approach: a systematic review and frequentist network meta-analysis,” Dec. 01, 2024, BioMed Central Ltd. doi: 10.1186/s13018-024-04852-3. [9] F. Moshfeghifar, K. Erleben, and S. Darkner, “PhD Thesis Hip Joint Finite Element Modeling.” [10] M. Isla, “Hip Joint: A Significant Component of Body Movement and Stability,” 2024, doi: 10.35841/2572-4916.24.12.304. [11] N. Sandiford, D. Kendoff, and S. Muirhead-Allwood, “Osteoarthritis of the hip: aetiology, pathophysiology and current aspects of management,” Jan. 01, 2020, AME Publishing Company. doi: 10.21037/aoj.2019.10.06. [12] C. Furlan et al., “Quality Of Life In Patients Submitted To Arthroplasty Total Hip Resurfacing,” Int. Arch. Med., 2015, doi: 10.3823/1742. [13] R. Condell, M. Kelly, and P. Mckenna, “Anatomic Markers for Intraoperative Identification of the Heuter Interval in the Direct Anterior Approach to the Hip,” 2023. [Online]. Available: www.techortho.com [14] D. A. Flevas, A. G. Tsantes, and A. F. Mavrogenis, “Direct Anterior Approach Total Hip Arthroplasty Revisited,” Apr. 15, 2020, Journal of Bone and Joint Surgery Inc. doi: 10.2106/JBJS.RVW.19.00144. [15] “Traction Table-related Complications in Orthopaedic Surgery,” 2010. [16] J. Nakamura et al., “Direct anterior approach for total hip arthroplasty with a novel mobile traction table -a prospective cohort study,” BMC Musculoskelet. Disord., vol. 18, no. 1, pp. 1–9, Jan. 2017, doi: 10.1186/s12891-017-1427-2. [17] A. Aprato, G. Parisi, A. D’amelio, and A. Massè, “Postless Hip Traction Table for Combined Femoral and Pelvic Fractures,” 2021. [Online]. Available: www.techortho.com|205 [18] R. Zhang, Y. Yin, S. Li, L. Jin, Z. Hou, and Y. Zhang, “Traction table versus double reverse traction repositor in the treatment of femoral shaft fractures,” Sci. Rep., vol. 8, no. 1, Dec. 2018, doi: 10.1038/s41598-018-24317-y. [19] T. Fathima, “Finite Element Analysis of Stress in Femur Bone,” 2020. [Online]. Available: www.ijprse.com [20] S. A. Maas, B. J. Ellis, G. A. Ateshian, and J. A. Weiss, “FEBio: Finite elements for biomechanics,” J. Biomech. Eng., vol. 134, no. 1, 2012, doi: 10.1115/1.4005694. [21] N. A. DeVries, K. H. Shivanna, S. C. Tadepalli, V. A. Magnotta, and N. M. Grosland, “Ia-FEMesh: anatomic FE models--a check of mesh accuracy and validity,” Iowa Orthop J, vol. 29, pp. 48–54, 2009. [22] A. Y. Bavil et al., “Effect of different constraining boundary conditions on simulated femoral stresses and strains during gait,” Sci. Rep., vol. 14, no. 1, Dec. 2024, doi: 10.1038/s41598-024-61305-x. [23] A. D. Speirs, M. O. Heller, G. N. Duda, and W. R. Taylor, “Physiologically based boundary conditions in finite element modelling,” J. Biomech., vol. 40, no. 10, pp. 2318–2323, 2007, doi: 10.1016/j.jbiomech.2006.10.038. [24] N. Bn and C. Bendigeri, “Finite element material models for nonlinear analysis of cortical bone-A Systematic Review,” 2021. [Online]. Available: www.ijcrt.org [25] M. Baleani, P. Erani, A. Acciaioli, and E. Schileo, “Tensile Yield Strain of Human Cortical Bone from the Femoral Diaphysis Is Constant among Healthy Adults and across the Anatomical Quadrants,” Bioengineering, vol. 11, no. 4, Apr. 2024, doi: 10.3390/bioengineering11040395. [26] M. Bittner-Frank, A. G. Reisinger, O. G. Andriotis, D. H. Pahr, and P. J. Thurner, “Cortical and trabecular mechanical properties in the femoral neck vary differently with changes in bone mineral density,” JBMR Plus, vol. 8, no. 6, Jun. 2024, doi: 10.1093/jbmrpl/ziae049. [27] T. Lim and J. Hong, “Poroelastic properties of bovine vertebral trabecular bone,” J Orthop Res, vol. 18, no. 4, 2000. [28] F. Bini, A. Marinozzi, F. Marinozzi, and F. Patanè, “Microtensile measurements of single trabeculae stiffness in human femur,” J Biomech, vol. 35, no. 11, Nov. 2002, Accessed: Apr. 16, 2026. [Online]. Available: 10.1016/s0021-9290(02)00182-3 [29] R. W. Ingle, “Finite element modelling the natural human joint,” 2009. [30] H. Jin and J. L. Lewis, “Determination of Poisson’s Ratio of Articular Cartilage by Indentation Using Different-Sized Indenters,” J. Biomech. Eng., vol. 126, no. 2, pp. 138–145, Apr. 2004, doi: 10.1115/1.1688772. [31] F. Moshfeghifar et al., “LibHip: An open-access hip joint model repository suitable for finite element method simulation,” Comput. Methods Programs Biomed., vol. 226, Nov. 2022, doi: 10.1016/j.cmpb.2022.107140. [32] M. D. Harris, A. E. Anderson, C. R. Henak, B. J. Ellis, C. L. Peters, and J. A. Weiss, “Finite element prediction of cartilage contact stresses in normal human hips,” Journal of Orthopaedic Research, vol. 30, no. 7, pp. 1133–1139, Jul. 2012, doi: 10.1002/jor.22040. [33] K. Trachoo, I. Chaiya, and D. Prathumwan, “Finite Element Analysis of Stress and Displacement in the Distal Femur: A Comparative Study of Normal and Osteoarthritic Bone Under Knee Flexion,” Computation, vol. 14, no. 1, p. 18, Jan. 2026, doi: 10.3390/computation14010018. [34] P. Chytla and W. Wojnarowska, “Finite Element Analysis Of The Mechanical Behavior Of The Human Femoral Bone: A Comparative Study Of Homogeneous Isotropic and Heterogeneous Isotropic Material Models,” 2025. [35] F. Castillo, C. A. Hernández-Salazar, and O. A. González-Estrada, “Structural analysis of the hip joint using segmentation and finite elements in patients with femoroacetabular impingement,” Front. Mech. Eng., vol. 11, 2025, doi: 10.3389/fmech.2025.1531864. [36] D. S. Cronin, B. Watson, F. Khor, D. Gierczycka, and S. Malcolm, “Cortical bone continuum damage mechanics constitutive model with stress triaxiality criterion to predict fracture initiation and pattern,” Front. Bioeng. Biotechnol., vol. 10, Oct. 2022, doi: 10.3389/fbioe.2022.1022506. [37] M. A. Röling, N. M. C. Mathijssen, I. Blom, T. Lagrand, D. Minderman, and R. M. Bloem, “Traction force for peroperative hip dislocation in hip arthroscopy,” HIP International, vol. 30, no. 3, pp. 333–338, May 2020, doi: 10.1177/1120700019841579. [38] E. F. Morgan, G. U. Unnikrisnan, and A. I. Hussein, “Bone Mechanical Properties in Healthy and Diseased States,” Jun. 04, 2018, Annual Reviews Inc. doi: 10.1146/annurev-bioeng-062117-121139. [39] N. H. Hart, S. Nimphius, T. Rantalainen, A. Ireland, A. Siafarikas, and R. U. Newton, “Mechanical basis of bone strength: influence of bone material, bone structure and muscle action,” 2017. [Online]. Available: http://www.ismni.org [40] J. A. Izquierdo, R. Diaz, C. R. Luna Lizarraga, K. Sclar, and J. Camacho-Galindo, “Surgical Approach Influences the Risk of Intraoperative Periprosthetic Femoral Fracture in Primary Total Hip Arthroplasty: A Retrospective Cohort Study,” Cureus, Mar. 2026, doi: 10.7759/cureus.106065.
local.eprints.dirdisk0/00/53/33/60
local.eprints.division07E
local.eprints.id533360
local.eprints.typethesis/2Sarjana
local.eprints.urihttp://scholar.unand.ac.id/533360/
local.fakultasFakultas Teknik
local.identifier.nim2210913038
local.prodiS1 Teknik Mesin
local.thesis.levelSkripsi

Files

Collections